Contact an Account ManagerPesticides: Organophosphorus compounds from water
Summary
Type: SPE
Application #: 301580
Catalog#: 730108 - SPE Cartridges, Chromabond, HR-P, 3ml, PP, 200mg 30/pk
Detection:
Conditions: Column conditioning:
2 x 2 ml ethyl acetate, 3 x 2 ml methanol, then 3 x 2 ml dist. water
Sample application:
Aspirate adjusted water sample through the column
Column washing:
2 ml dist. water, then dry column for 30 min under vacuum
Elution:
2 x 1 ml ethyl acetate / n-hexane (50:50, v/v)
Authors: Macherey-Nagel
Source: LC Department
Details
Legend: Recovery of pesticides from water: (GC) , 150 ng/l per component
Compound recovery [%]
Azinphon methyl 90
Carbophenothion 89
Chlorphenvinphos 88
Demeton-s-methyl 91
Diazinon 95
Dichlorphos 90